Current clamp PCE-LCT 3 

Multimeter for determining the leakage current / Temperature measurement with thermocouple / Inductive current measurement up to 150 A AC / Current probe opening 30 mm / Many measuring functions

The multimeter is the ideal measuring instrument to measure leakage currents. This is made possible by the extensive measuring range of the multimeter. Also, current measurements up to 150 A AC can be made with the multimeter. To test supply lines with a current of up to 150 A, the multimeter for the corresponding cable cross-section can be opened up to 30 mm. 
In addition to the current measurement, the multimeter can also measure voltages up to 600V directly. Because of the current and voltage measurement, the multimeter is the perfect measuring instrument to carry out measurements and analyzes on, for example, three-phase networks. For example, the continuity test can be used to find a wrong wiring in sub-distribution boxes.

– 30 mm current clamp opening
– Temperature measurement with thermocouple type K
– Current measurement up to 150 A AC
– Covers a wide range of measurements
– Voltage measurement up to 600V AC / DC RMS
– Transport bag for mobile use



Specification

AC current measurement  
measuring range Resolution    Accuracy Accuracy
4-mA 0.001-mA ± (2% + 10 digits) ± (3% + 5 digits)
40-mA 0.01-mA ± (2% + 10 digits) ± (3% + 5 digits)
400-mA 0.1-mA ± (2% + 5 digits) ± (3% + 3 digits)
4 A 0.001 A ± (2% + 5 digits) ± (3% + 3 digits)
40 A 0.01A ± (2% + 10 digits) ± (3% + 5 digits)
150 A 0.1 A ± (2% + 10 digits) ± (3% + 5 digits)
       
Frequency range: 40 Hz … 1 kHz (sine wave)
Maximum input current: 150 A AC
       
 
DC voltage  
measuring range Resolution Accuracy
4V 0.001V ± (0.5% + 4 digits)
40V 0.01V  
400V 0.1V  
600V 1V  
     
Input impedance: 10 MΩ
Maximum input voltage: 600V AC / DC RMS
     
 
AC voltage   
measuring range Resolution Accuracy
4V 0.001V ± (1% + 3 digits)
40V 0.01V  
400V 0.1V  
600V 1V  
     
Input impedance: 10 MΩ
maximum input voltage: 600V AC / DC RMS
Frequency range: 40 Hz … 1 kHz (sine wave)
     
 
Resistance  
measuring range Resolution Accuracy
400 Ω 0.1 Ω ± (0.8% + 3 digits)
4 kΩ 0.001 kΩ  
40 kΩ 0.01 kΩ  
400 kΩ 0.1 kΩ  
4 MΩ 0.001 MΩ  
40 MΩ 0.01 MΩ ± (1% + 3 digits)
     
Measuring voltage in open circuit: approx. 0.4V
Overvoltage protection: 600V AC / DC RMS
     
 
Continuity test  
measuring range Resolution Accuracy
Beep at < 40 Ω 0,1Ω
     
Overvoltage protection: 600V AC / DC RMS
     
 
Temperature  
measuring range Resolution Accuracy
-20 … 0°C 1°C ± (3% + 5 digits)
-4 … 32°F 1°F  
-0 … 400°C 1°C ± (1.5% + 5 digits)
32 … 757°F 1°F  
400 … 1000°C 1°C ± (3% + 5 digits)
752 … 1832°F 1°F  
     
Overvoltage protection: 600V AC / DC RMS
The temperature accuracies do not include the accuracies of the temperature sensors
     
 
Capacity  
measuring range Resolution Accuracy
40.00 n 0.01 nF ± (3% + 8 digits)
400.0 nF 0.1 nF  
4,000 μF 0.001 μF  
40.00 μF 0.01 μF  
400.0 μF 0.1 μF  
4,000mF 0.001 μF  
40.00 mF 0.01 μF  
     
Overvoltage protection: 600V AC / DC RMS
     
 
Diode test  
measuring range Resolution Accuracy
3.2V 0.001V
     
Overvoltage protection: 600V AC / DC RMS
Test voltage: 3.2V
Forward current: 1-mA
The accuracies refer to the following ambient conditions: 23 ± 5°C / 9°F and < 75% RH
   
General specifications  
Maximum current clamp opening 30 mm
Automatic shutdown After 30 minutes, can be deactivated
Power supply 2 x 1.5V AAA battery
Measuring rate 3 measurements per second
Display with overvoltage “OIL”
Maximum display 4000
Display LCD display
Maximum working height 2,000 m / 13,123 ft
Temperature coefficient 0.1 x accuracy x °C, °F*
Operating conditions 18 … 28°C / 64 … 82°F
Store control admixtures -10 … 50°C / 14 … 122°F
Dimensions 213 x 62 x 38 mm / 8.4 x 2.4 x 1.5 in
Weight About 238 g / < 1 lb with batteries
   
*The temperature is the difference between the temperature of the operating conditions and the current ambient temperature.
 
Example:
Is the current ambient temperature greater than the temperature of the operating conditions
(50°C / 122°F (current ambient temperature)) – (40°C 104°F (operating temperature)) = 10°C / 50°F
Is the current ambient temperature less than the temperature of the operating conditions
(0°C / 32°F (operating temperature)) – (-5°C / 23°F (current ambient temperature)) = 5°C / 41°F
 



Delivery Scope

1 x Current clamp PCE-LCT 3
1 x Transport bag
1 x Thermocouple type K
1 x Pair of test leads
2 x 1.5V AAA batteries
1 x User manual



Downloads